Series 5760 OEM material detector


 "Teachable" Detector Based on True False Principal


5 7 6 0                  
   
 

Options: 

1 mm2 surface detection area
    Line detecting for large surfaces
   
 
 
 

 

   
3D AutoCAD drawing of sensor (zipped)   -   DWG = [R13] [R14] [R2000]       DXF = [R13] [R14] [R2000]

Technical Data Connections
Operating temp: - 20 ...+   60 degree C 4 pole square connector
- 04 ...+ 140 degree F GND = GND
Weight:    6 oz (0.17 kg) 1 = Power Supply +11 ... 24 Volts, 100 mA consumption
Protection: IP 54, NEMA 3 2 = Push Pull output, High / Low, max 50 mA
3 = Push Pull output, High in tolerance field, max 50 mA
Housing: Die Cast Zinc Alloy
Humidity: Up to 98% permissible
Spectral Range: 620 ... 680 nm
Capture time:  30 msec
Data capture:  6 x 10 bit
Description  
The polarimeter analyzes the polarization state of light. Application software shows degree-of-polarization, azimuth angle, ellipticity, normalized Stokes vector and Jones vector. Device’s coordinate frame is shown in the instruction file. Input aperture has a diameter of 3mm. Due to sensitivity it is recommended that the polarimeter be mounted firmly and light under analysis must fall straight in to the device. If it is possible, speckles in light should be avoided. If the device is included in an ellipsometer, it calculates surface’s complex refractive index.

First, the detector is placed over a master sample and the button is pressed and held for three seconds.  The data is then stored or "taught" to the device.  Second, immediately  after the data has been stored, depress the same button again in short presses to set the tolerance.  Each press is equivalent to 0.2% tolerance.  The maximum tolerance is 3.2%.  There are 20 seconds to "calibrate" the detector.

Example:  To set the detector to sense a a certain polarization state from a  master with a 1.2% tolerance.  The detector is placed on the sheet and the button is pressed for 3 seconds and released, then the button is pressed in short increments 6 times. (6 x 0.2 = 1.2)

Output
Sample of Output
Part 1: Low out of Tolerance
Part 2: Low in Tolerance
Part 3: Hi in Tolerance
Part 4: Part 4: Hi out of Tolerance
Dimensions
For a nominal charge of $100.00, Hohner Corp. will test and evaluate your samples to ensure this product meets your specific requirements and standards.  Should modifications be required, Hohner Corp. will advise in writing of any additional charges.

Cleaning of Device: Should be done regularly with a clean, dry, lint free cotton cloth.

The Life expectance of this device is 100,000 hours